Automated ELISA processors (BIOBASE4000)

Product#: DCBio-0004
$57,000.00
Availability:
Ships in 1-2 weeks

Automated ELISA processors (BIOBASE4000)


Cat. No. DCBio-0004

Description

BIOBASE 4000 ELISA processors is fully autoated Four-plates ELISA processing system, capable of sampling, incubation, washing and data processing, all automatically.

It is complete solution for your IVD ELISA assays. The system combines a high precision pipetting unit with stable incubations and efficiency washers and photometer unit. 


Features

  1. 1 robotic arm, 4 pipetting probe(10~1000μl). 
  2. 8 unit 96 well microplates(independent incubating). 
  3. 2 unit reader, 2 unit washer (auto reading and washing).

01.png

Software Module

  • User-friendly Windows system
  • LIS system available

Sample Module

Hitech teflon coating probe to prevent cross contamination.

Sample& Reagent& Dilution Rack Module

Original sample tubes available;Rack positions programmable.

Microplate washer& Reader module

Modularized automatic control reading and washing system.

Models

Model

DCBio-0004

 

 

Sample Unit

Sample Racks

Standard 600 sample positions

Sample Tubes

11.5~13mm

Pipetting

8~1000ul

Dispensing system

4 aspirating-dispensing probe (X-Y-Z movement)

Reagent Unit

 

Reagent Position

23 reagent racks, editable

Pipetting Range

8~1000ul, 1ul stepping

Pipetting Precision

(100ul) CV≤0.5%

Dispensing Time

3 minutes to whole96well microplates

Washing Probes

Two line 8-nozzle manifolds

Wash Containers

1Cleaning fluid 15L, 1waste water 25L, 1Buffer 5L and 3 Buffer 2L, with liquid level sensor, liquid shortage and full alarm

Washing Residual

<2μl

Reading Unit

 

Reading Channels

8 independent photometric channels, mono

and dichromatic reading

Absorbance

0.0~3.0 OD

Spectral Range

400-700nm

Optical Filters

2standard filters (450,630nm);

2 more on request (405nm/450nm/492nm/630nm)

Reading Accuracy

±3% from 0.2~1.0 OD, ±2% from 1.0 to1.5 OD

Incubation Unit

Incubation

8 independent incubators

Temperature Range

RT to 45 °C

Temperature Accuracy

±0.5°C

Data Management

Software

Windows 7

LIS System

Bi-direction, support HL7 protocol

Results

Absorbance and results reviewable by software

Printer

External optional

Working Conditions

Power Supply

AC220V±10%,50/60Hz, 110V±10%,60Hz, 800W

Temperature

1030℃

Humidity

30-80%

Size & Weight

External Size (W*D*H)

1343*860*1700mm

Net Weight

215kg

Package Size (W*D*H)

Machine:1520*970*1108mm

Base Stand:1520*1020*1378mm

Gross Weight

Machine:220kg

Base Stand: 165kg

Diagnocine-Title Logo