3 Part Auto Hematology Analyzer (BK-9610)

Product#: DCBio-0916-KB
$6,600.00
Availability:
Ships in 1-2 weeks

Auto Hematology Analyzer (BK-9610)


Cat. No. DCBio-0916-KB

Features

  1. Throughput: 60T/H
  2. 8 inch touch screen
  3. 20 parameters + 3 histograms
  4. 3 counting modes
  5. 100,000 sample results
  6. Support LIS and external printer 

Model
 

Model

DCBio-0916-KB

Throughput

60Tests/hour

Measuring Principle

Impedance for cell counting Cyanide-free method for HGB

Aperture Diameter

WBC chamber:100μm, RBC/PLT chamber:70μm

Parameters

WBC, Neu#, Lym#, Mid#, Neu%, Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR

Histograms

3 histograms (including WBC, RBC, PLT histogram)

Sample Volume

Normal:10μl; Predileted:20μl

Calibration

Manual & auto calibration

Counting Modes

Venous, capilary, prediluted

Printer

Built-in thermal printer, USB to external printer(optional)

Data Input

8-inch touch screen, keyboard, mouse

Output

RS232 port, USB port, VGA port. LIS system available

Storage

100,000 results with histogram

Carry-over rate

Parameter

CV

WBC

≤ 2%

RBC

≤ 1%

HGB

≤ 2%

PLT

≤ 2%

Background

Parameter

Background

WBC

≤ 0.2×10^9/L

RBC

≤ 0.02×10^12/L

HGB

≤ 1 g/L

HCT

≤ 0.5%

PLT

≤ 10×10^9/L

 

Precision

 

Parameter

CV

Measurement range

WBC

≤ 2.0%

(4.0-15.0×10^9)/L

RBC

≤ 1.9%

(3.5-6.0×10^12)/L

HGB

≤ 1.9%

(110.0-180.0g)/L

MCV

≤ 0.4%

(80.0-110.0)fL

PLT

≤ 4.0%

(100.0-500.0×10^9)/L

Linearity

Parameter

Measurement range

CV

WBC

(0-5.9×10^9)/L

(6.0-99.9×10^9)/L

(±0.3×10^9)/L

±5%

RBC

(0-0.99×10^12)/L

(1.0-9.99×10^12)/L

(±0.05×10^12)/L

±5%

HGB

(0-99.0g)/L

(100.0-300.0g)/L

(±2.0g)/L

±2%

PLT

(0-99×10^9)/L

100-999×10^9)/L

(±8.0×10^9)/L

±10%

Power supply

AC 220±10%,60/50Hz, 110±10%,60Hz

Package Size

580*470*590mm (W*D*H)

Reagent Package Size

450*410*310mm (W*D*H)

Gross Weight

27kg

Reagent Gross Weight

23kg

Diagnocine-Title Logo.jpg

Satisfaction
Quality Rating
Value Rating
Style Rating